SIRT / TARE Dosimetry Calculator

SIRT / TARE Dosimetry Calculator

Y-90 radioembolisation — MIRD (mono-compartment), BSA, and Partition models for HCC and liver metastases.

⚠ NOT FOR CLINICAL USE. This tool is an independent estimate for personal cross-checking and record-keeping only. It must not be used to determine, verify, or alter patient treatment. Always rely on validated, calibrated clinical dosimetry software and local governance.

Patient (optional)

Volumes & Physiology

Advanced assumptions

Dose constant: 49.67 Gy·kg/GBq (Y-90, full decay). Mass = volume × density.

Microspheres & FLEX

Prescribed Activity

1.50 GBq
02.557.510 GBq
Estimated microspheres million

Calculated Doses

MIRD — mono-compartment whole perfused volume

Treatment area (mean)uniform-distribution assumption Gy
Whole-liver equivalentif activity spread over total liver Gy

Partition model uses T:N

Tumour dosewithin treatment area Gy
Normal liver (treated)non-tumour, perfused Gy

Lung shared

Lung dosevia shunt fraction Gy

BSA method resin · recommended activity

Recommended activityDu Bois BSA GBq
Implied mean treated doseat recommended activity Gy

Green ≤70 Gy / orange >70 Gy on treated normal liver. Lung: yellow >15 Gy, red >30 Gy (IFU single-administration limit 25 Gy). FLR <30% flagged red.

Liver Diagram

Vessel width
Vessel Admin point (★) Tumour (ovoid)

Vessel: click-drag to draw freehand. Admin point: click to drop a star. Tumour: click-drag to size an ovoid. Liver image: load your own picture as a faded backdrop (adjust the Faint slider), then draw over it.

Export & Reset

PDF/JPEG include patient details, all three model results, and the annotated diagram, with the not-for-clinical-use notice.

Spectrum IR — internal cross-check tool. Constants & formulas per published Y-90 SIRT dosimetry literature.